How to do the calibration on ADAM-4013? | |||
---|---|---|---|
Solution:
Please do the following steps for calibration :
1. Execute ADAM.EXE
2. Search the ADAM-4013 module
3. Select Calibration item
4. Wiring should be
R Connection
+ +IEXC, +SENSE
- -IEXC, -SENSE, AGND
Note : The AGND is ground reference that needs to be connected when applying IEXC for exciting power source. It also provide a reference ground for external power as exciting source.
5. Do offset calibration R = 60 Ohm, and span calibration R = 140 Ohm
6. Connect a variable VR(200 Ohm) for testing
|