Doc. # 1-0003224 | |||
---|---|---|---|
Date Updated | 01-22-2004 | Date Created | 07-27-2000 |
Document Type | Knowledge Base | Related OS | |
Related Product | ADAM-4013 |
ADAM-4013 accuracy specification change note. | |||
---|---|---|---|
Solution:
The zero drift and span drift specifications of ADAM-4013 in the annual catalog and the manual are incorrect. The correct value for both drift specifications is 0.01_ C/_ C. This means the measured temperature will increase or decrease 0.01_ C when environmental temperature is changed by 1 _ C.
|